RTD resistance/temperature calculator

Uses the IEC 60751 curve (α = 0.00385). Changing the temperature unit converts the entered temperature.

Enter the measured resistance to see the temperature.

No valid result to draw.

Show the steps

Interactive calculation is unavailable until JavaScript loads. The formulas, worked examples and references below remain available.

Convert Pt100 or Pt1000 RTD resistance to temperature and back with the IEC 60751 Callendar-Van Dusen equation from -200 to 850 °C.

How to use

Choose Temperature to resistance or Resistance to temperature.

Choose Pt100, Pt1000 or another R0.

Enter the temperature with its unit, or the measured resistance in ohms.

For a 2-wire probe, enter the resistance of each lead; leave 0 for 3- or 4-wire wiring.

Read the result, its sensitivity and the straight-line comparison. Copy result copies labeled values; Reset restores the starting fields.

Science

A platinum RTD's resistance rises almost linearly with temperature: a Pt100 is 100 Ω at 0 °C and about 138.5 Ω at 100 °C. IEC 60751 defines the curve with the Callendar-Van Dusen equation and three coefficients, so any Pt100 or Pt1000 to the standard can be converted with the same formula.

Above 0 °C the equation is a quadratic, which can be solved exactly for temperature. Below 0 °C a fourth-order term is added, so the tool solves it by Newton iteration, starting from the quadratic answer and stopping when the step is below 1e-10 °C.

The common shortcut R = R0 × (1 + 0.00385 × t) is exact only at 0 and 100 °C. At 400 °C it overstates a Pt100 by 6.93 Ω, an 18 °C error. With 2-wire wiring the lead resistance adds to the reading: 0.5 Ω per lead makes 100 °C read as 102.6 °C, so 3- and 4-wire connections are used to cancel it.

Limitations: the tool gives the ideal IEC 60751 curve. A real sensor differs within its tolerance class (not computed here), and self-heating, lead resistance changes with temperature, insulation leakage and the measuring instrument add error. Results are for -200 to 850 °C only, with the α = 0.00385 curve; other curves and ADC codes are not included.

Formulas

At or above 0 °C: R = R0 × (1 + A × t + B × t^2).

Below 0 °C: R = R0 × (1 + A × t + B × t^2 + C × (t - 100) × t^3).

Coefficients: A = 3.9083e-3 /°C, B = -5.775e-7 /°C^2, C = -4.183e-12 /°C^4 (IEC 60751).

Inverse at or above 0 °C: t = (-A + sqrt(A^2 - 4 × B × (1 - R/R0))) / (2 × B).

Inverse below 0 °C: Newton: t → t - f(t)/f'(t), f(t) = R(t) - R, from the quadratic estimate, until the step is below 1e-10 °C.

Alpha and leads: α = (R100 - R0)/(100 × R0) = 0.00385055; 2-wire reading = R + 2 × R_lead.

Examples

Pt100 at 100 °C: 138.5055 Ω. 100 × (1 + 0.39083 - 0.005775); tables print 138.51 or 138.505.

Pt100 reading 110 Ω: 25.684 °C. Closed-form quadratic.

Pt100 reading 60.2558 Ω: -100.000 °C. Below 0 °C: Newton iteration, 3 steps. The quadratic alone would give -100.208 °C.

Pt1000 at -200 °C and 850 °C: 185.2008 Ω and 3904.8113 Ω. Boundary: the ends of the IEC range; Pt1000 is 10 × Pt100.

Common mistake: straight line at 400 °C: 254.022 Ω instead of 247.092 Ω. 0.00385 per °C is the 0 to 100 °C average, not the slope everywhere.

Pt100 at 100 °C, 2-wire, 0.5 Ω per lead: meter reads 139.5055 Ω, which is 102.638 °C if the leads are ignored. Subtract both leads, or use 3- or 4-wire wiring.

References

IEC 60751:2022 (IEC webstore)

TI SBAA275A: A basic guide to RTD measurements

Microchip AN687: Precision temperature-sensing with RTD circuits

TI analog engineer's pocket reference: RTD equations

Minco: RTD temperature vs. resistance table (100 Ω)

BIPM: SI Brochure