Strain-gauge/load-cell bridge calculator

Positive strain is tension. Output is positive when the +ε gauges stretch. The load unit only relabels the capacity and load.

Enter the gauge factor, strain and excitation to see the bridge output.

No valid result to draw.

Show the steps

Interactive calculation is unavailable until JavaScript loads. The formulas, worked examples and references below remain available.

Find quarter-, half- or full-bridge output from gauge factor and strain, strain from a measured output, or a load cell's output at a load from its rated output in mV/V.

How to use

Choose a task: strain to output, output to strain, load to output or output to load.

For strain, choose the bridge arrangement and enter the gauge factor from the gauge packet.

Enter the excitation voltage and the strain in microstrain (negative for compression), or the measured output in millivolts.

For a load cell, enter the rated output in mV/V, the rated capacity and unit, and the load or the measured output; enter the no-load output if the cell is not zeroed.

Read the output, strain or load and the secondary values. Copy result copies labeled values; Reset restores the starting fields.

Science

A strain gauge's resistance changes in proportion to strain: ΔR/R is the gauge factor times the strain, about 2 for foil gauges. Because the change is tiny (0.2 % at 1000 με), gauges are read in a Wheatstone bridge, which turns it into a small voltage proportional to the excitation, quoted in mV/V.

With one active gauge (a quarter bridge) the output is not exactly linear: the gauge's own change unbalances the current in its arm. The error is about the gauge factor times the strain divided by two, about 0.1 % at 1000 με and 1 % at 10,000 με for a gauge factor of 2. Two gauges with equal and opposite strain in adjacent arms (bending), or four in opposite pairs, cancel that term and are exactly linear, with two and four times the output.

A load cell is a full bridge inside a calibrated body. Its rated output in mV/V is the output change from zero to rated load with 1 V excitation, so a 2 mV/V cell at 5 V gives 10 mV at full load and the output scales in proportion to the load.

Limitations: the gauges and completion resistors are ideal and equal, lead resistance is zero, and the excitation is constant. Poisson-gauge arrangements, lead desensitisation, temperature effects, the load cell's own nonlinearity and hysteresis, and amplifier or ADC scaling are not included. The sign shown follows the stated arm layout, as Vishay and HBK print it; NI's equations use the opposite sign, and swapping the output leads reverses it.

Formulas

Gauge factor: GF = (ΔR/R) / ε. ε in m/m; 1 με = 1e-6.

Quarter bridge: Vout / Vex = GF × ε / (4 + 2 × GF × ε).

Quarter-bridge inverse: ε = 4 × Vr / (GF × (1 - 2 × Vr)), Vr = Vout / Vex.

Half and full bridges: Vout / Vex = GF × ε / 2 (half); GF × ε (full).

Nonlinearity: η = GF × ε / (2 + GF × ε); exact = linear × (1 - η).

Load cell: Vout = V0 + C × Vex × F / Fcap. C in mV/V.

Examples

Quarter bridge, GF 2, +1000 με, 5 V: 2.4975 mV (0.4995 mV/V). Linear estimate 2.5 mV is 0.0999 % high.

Half and full bridges, same strain: 5 mV and 10 mV. Two and four active gauges with opposite strains; exactly linear.

Quarter bridge, GF 2, output 37.5 mV at 5 V: 15,228.4 με. Common mistake: the linear formula 4Vr/GF gives 15,000 με, 1.5 % low (Vishay TN-507's example).

Load cell 2 mV/V, 50 kg, 5 V, 20 kg: 4 mV of a 10 mV full scale. 2 × 5 × 20 / 50.

References

NI: Measuring strain with strain gages

NI-DAQmx: quarter-bridge type I

Vishay Micro-Measurements TN-507-1: Wheatstone bridge nonlinearity

HBK: The Wheatstone bridge circuit explained

HBK: Adjustment of electronics for force measurements

Analog Devices CN-0216: precision weigh scale design