Frequency-counter resolution calculator

Enter a decimal or scientific notation, such as 10 or 1e7. Changing a unit reinterprets its number.

One period of the counting clock (100 ns for a 10 MHz clock), or the single-shot resolution from the data sheet.

The timebase specification, for example 1 for a ±1 ppm crystal oscillator.

Enter the input frequency to see the resolution.

QuantityConventionalReciprocal

No valid result to draw.

Show the steps

Interactive calculation is unavailable until JavaScript loads. The formulas, worked examples and references below remain available.

Compare the ±1-count resolution of conventional and reciprocal frequency counters for a gate time, add timebase error, or find the gate time for a target resolution.

How to use

Choose Resolution for a gate time or Gate time for a resolution, and the counter type to highlight.

Enter the input frequency and the gate time, or the target resolution.

Enter the counter's time resolution and, optionally, its timebase uncertainty in ppm.

Read the resolution or gate time, and compare both counter types in the table and chart.

Copy result copies labelled values; Reset restores the starting fields.

Science

A conventional counter opens a gate for a fixed time and counts input cycles. The gate is not synchronised to the input, so the count can be one more or one less: the resolution is 1 divided by the gate time, 1 Hz for a 1 s gate, whatever the frequency. That is fine at 10 MHz but only 3 digits at 1 kHz.

A reciprocal counter waits for whole input periods and times them with its own clock, then divides. The ±1 count is now a clock period, so the relative resolution is the clock period divided by the measurement time, the same at every frequency. Below the clock frequency it resolves better than direct counting.

Resolution is not accuracy. The timebase oscillator's error, in ppm, shifts every reading by the same fraction: 1 ppm is 10 Hz at 10 MHz. Once it is larger than the resolution, a longer gate adds digits but not accuracy.

Limitations: the tool models the ±1-count resolution and adds the timebase error as a worst-case sum. Trigger error from noise and slow edges, systematic errors and the resolution enhancement and averaging of modern counters (specified in digits per second) are not modelled, and the timebase uncertainty you enter should include aging and temperature.

Formulas

Conventional resolution: Δf = ±1 / T_g. T_g is the gate time.

Conventional relative resolution: Δf / f = 1 / (f × T_g).

Reciprocal relative resolution: Δf / f = t_res / T_m, T_m = n / f with n = ceil(f × T_g) whole input periods.

Digits: D = log10(f / Δf).

Timebase error: Δf_tb = f × u × 10^-6, u in ppm.

Gate time: T_g = 1 / Δf (conventional); T_g = f × t_res / Δf, at least one period (reciprocal).

Examples

1 kHz, 1 s gate, 100 ns time resolution: Reciprocal ±0.1 mHz (7 digits); conventional ±1 Hz (3 digits). 100 ns / 1 s = 1 × 10^-7 of 1 kHz.

10 MHz, 1 s gate, 100 ns: Both ±1 Hz (7 digits). The counters are equal at f = 1 / 100 ns = 10 MHz.

100 MHz, 0.1 s gate, 100 ns: Conventional ±10 Hz; reciprocal ±100 Hz. Above the clock frequency direct counting is finer.

10 MHz, 1 s gate, 1 ppm timebase: Resolution ±1 Hz; timebase ±10 Hz; worst case ±11 Hz. AN 200's example: the timebase dominates.

50 Hz, 1 s gate, conventional: ±1 Hz, 2 % of the reading. Common mistake: direct counting at low frequency. A reciprocal counter gives ±5 μHz.

1 kHz, target 1 mHz, 100 ns: Conventional 1000 s; reciprocal 0.1 s. 1 / 0.001 Hz against 1000 × 100 ns / 0.001 Hz.

0.5 Hz, 1 s gate: Conventional reads 0 or 1 count (±1 Hz); reciprocal times one 2 s period (±25 nHz). Boundary: fewer than one cycle in the gate.

References

HP Application Note 200: Fundamentals of the electronic counters

HP Application Note 200-4: Understanding frequency counter specifications

NIST: Frequency measurement (Lombardi)

NIST SP 1500-203: Timing annex

Keysight 53200A series counter data sheet