HONEYWELL XCX15DNC Board Mount Pressure Sensor - 15 psi Unamplified
Introducing the Honeywell XCX15DNC, a high-precision silicon micromachined pressure sensor designed for accurate pressure measurement in various industrial and medical applications. Manufactured in the USA, this board-mounted transducer offers reliable performance with temperature compensation and calibration integrated into an industry-standard package.
Product Features
Pressure range up to 15 psi (1 psi to 240 psi options available)
Unamplified ratiometric millivolt (mV) output
Temperature compensated from 0 °C to 70 °C (32 °F to 158 °F)
Available in High Grade (H) for high-performance sensing or Commercial Grade (C) for lower performance needs
Protected against torque-induced stress with a unique stress isolating design
Calibrated offset with high long-term stability and accuracy
Suitable for gage, differential, and absolute pressure measurements
Electrical Specifications (at 12 Vdc excitation, 25 °C)
Characteristic
XCX/XCXL C Grade
XCX/XCXL H Grade
Unit
Excitation voltage
3.0 - 16.0 Vdc
3.0 - 16.0 Vdc
Null
±0.5 - ±1.0 mV
±0.5 - ±1.0 mV
Temperature error on offset (0 °C to 70 °C)
±0.5 mV
±0.5 mV
Temperature error on span (0 °C to 70 °C)
1.0% span
1.0% span
Linearity, hysteresis error
0.5 - 1.0% span
0.3 - 0.5% span
Repeatability
0.1% span
0.1% span
Pressure Range and Output
Full Scale Pressure
Output Full Scale Span (mV)
Max Overpressure (psi)
4 in H2O
38 - 42
5
10 in H2O
19 - 21
5
0.3 psi
19 - 21
5
1 psi
17 - 19
5
5 psi
57 - 63
15
15 psi
85 - 95
45
30 psi
85 - 95
90
60 psi
85 - 95
180
100 psi
95 - 105
200
150 psi
85 - 95
300
240 psi
95 - 105
300
Absolute Output (at 12 Vdc, 25 °C)
FS Pressure
Output (mV)
Max Overpressure (psi)
5 psi
57 - 63
15
15 psi
85 - 95
45
30 psi
85 - 95
90
60 psi
85 - 95
180
100 psi
95 - 105
200
Specifications & Additional Information
Pressure ranges from 4 in H2O to 240 psi
Calibrated for full scale span with ±1.0% (H Grade) or ±2% (C Grade) accuracy over the temperature range
Suitable for medical, vacuum, positive pressure, and small size applications
RoHS compliant
ECCN: EAR99
Package Type: Board Mount
Datasheet: Download Datasheet
Disclaimers
Warning: Do not use these products as safety or emergency stop devices. Failure to follow installation and operation instructions may result in personal injury or death. Specifications are subject to change without notice. For detailed application and installation guidance, refer to the provided datasheet.
Part Number: XCX15DNC | Manufacturer: Honeywell | COO: US | UOM: EA | ECCN: EAR99 | Package Type: Board Mount
Use left/right arrows to navigate the slideshow or swipe left/right if using a mobile device
{"id":10028359090493,"title":"XCX15DNC","handle":"xcx15dnc","description":"\u003ch3\u003eHONEYWELL XCX15DNC Board Mount Pressure Sensor - 15 psi Unamplified\u003c\/h3\u003e\n\n\u003cp\u003eIntroducing the Honeywell XCX15DNC, a high-precision silicon micromachined pressure sensor designed for accurate pressure measurement in various industrial and medical applications. Manufactured in the USA, this board-mounted transducer offers reliable performance with temperature compensation and calibration integrated into an industry-standard package.\u003c\/p\u003e\n\n\u003ch3\u003eProduct Features\u003c\/h3\u003e\n\u003cul\u003e\n \u003cli\u003ePressure range up to 15 psi (1 psi to 240 psi options available)\u003c\/li\u003e\n \u003cli\u003eUnamplified ratiometric millivolt (mV) output\u003c\/li\u003e\n \u003cli\u003eTemperature compensated from 0 °C to 70 °C (32 °F to 158 °F)\u003c\/li\u003e\n \u003cli\u003eAvailable in High Grade (H) for high-performance sensing or Commercial Grade (C) for lower performance needs\u003c\/li\u003e\n \u003cli\u003eProtected against torque-induced stress with a unique stress isolating design\u003c\/li\u003e\n \u003cli\u003eCalibrated offset with high long-term stability and accuracy\u003c\/li\u003e\n \u003cli\u003eSuitable for gage, differential, and absolute pressure measurements\u003c\/li\u003e\n\u003c\/ul\u003e\n\n\u003ch3\u003eElectrical Specifications (at 12 Vdc excitation, 25 °C)\u003c\/h3\u003e\n\u003ctable\u003e\n \u003ctr\u003e\n \u003cth\u003eCharacteristic\u003c\/th\u003e\n \u003cth\u003eXCX\/XCXL C Grade\u003c\/th\u003e\n \u003cth\u003eXCX\/XCXL H Grade\u003c\/th\u003e\n \u003cth\u003eUnit\u003c\/th\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eExcitation voltage\u003c\/td\u003e\n \u003ctd\u003e3.0 - 16.0 Vdc\u003c\/td\u003e\n \u003ctd\u003e3.0 - 16.0 Vdc\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eNull\u003c\/td\u003e\n \u003ctd\u003e±0.5 - ±1.0 mV\u003c\/td\u003e\n \u003ctd\u003e±0.5 - ±1.0 mV\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eTemperature error on offset (0 °C to 70 °C)\u003c\/td\u003e\n \u003ctd\u003e±0.5 mV\u003c\/td\u003e\n \u003ctd\u003e±0.5 mV\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eTemperature error on span (0 °C to 70 °C)\u003c\/td\u003e\n \u003ctd\u003e1.0% span\u003c\/td\u003e\n \u003ctd\u003e1.0% span\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eLinearity, hysteresis error\u003c\/td\u003e\n \u003ctd\u003e0.5 - 1.0% span\u003c\/td\u003e\n \u003ctd\u003e0.3 - 0.5% span\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eRepeatability\u003c\/td\u003e\n \u003ctd\u003e0.1% span\u003c\/td\u003e\n \u003ctd\u003e0.1% span\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n\u003c\/table\u003e\n\n\u003ch3\u003ePressure Range and Output\u003c\/h3\u003e\n\u003ctable\u003e\n \u003ctr\u003e\n \u003cth\u003eFull Scale Pressure\u003c\/th\u003e\n \u003cth\u003eOutput Full Scale Span (mV)\u003c\/th\u003e\n \u003cth\u003eMax Overpressure (psi)\u003c\/th\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e4 in H2O\u003c\/td\u003e\n \u003ctd\u003e38 - 42\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e10 in H2O\u003c\/td\u003e\n \u003ctd\u003e19 - 21\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e0.3 psi\u003c\/td\u003e\n \u003ctd\u003e19 - 21\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e1 psi\u003c\/td\u003e\n \u003ctd\u003e17 - 19\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e5 psi\u003c\/td\u003e\n \u003ctd\u003e57 - 63\u003c\/td\u003e\n \u003ctd\u003e15\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e15 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e45\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e30 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e90\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e60 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e180\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e100 psi\u003c\/td\u003e\n \u003ctd\u003e95 - 105\u003c\/td\u003e\n \u003ctd\u003e200\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e150 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e300\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e240 psi\u003c\/td\u003e\n \u003ctd\u003e95 - 105\u003c\/td\u003e\n \u003ctd\u003e300\u003c\/td\u003e\n \u003c\/tr\u003e\n\u003c\/table\u003e\n\n\u003ch3\u003eAbsolute Output (at 12 Vdc, 25 °C)\u003c\/h3\u003e\n\u003ctable\u003e\n \u003ctr\u003e\n \u003cth\u003eFS Pressure\u003c\/th\u003e\n \u003cth\u003eOutput (mV)\u003c\/th\u003e\n \u003cth\u003eMax Overpressure (psi)\u003c\/th\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e5 psi\u003c\/td\u003e\n \u003ctd\u003e57 - 63\u003c\/td\u003e\n \u003ctd\u003e15\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e15 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e45\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e30 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e90\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e60 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e180\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e100 psi\u003c\/td\u003e\n \u003ctd\u003e95 - 105\u003c\/td\u003e\n \u003ctd\u003e200\u003c\/td\u003e\n \u003c\/tr\u003e\n\u003c\/table\u003e\n\n\u003ch3\u003eSpecifications \u0026amp; Additional Information\u003c\/h3\u003e\n\u003cul\u003e\n \u003cli\u003ePressure ranges from 4 in H2O to 240 psi\u003c\/li\u003e\n \u003cli\u003eCalibrated for full scale span with ±1.0% (H Grade) or ±2% (C Grade) accuracy over the temperature range\u003c\/li\u003e\n \u003cli\u003eSuitable for medical, vacuum, positive pressure, and small size applications\u003c\/li\u003e\n \u003cli\u003eRoHS compliant\u003c\/li\u003e\n \u003cli\u003eECCN: EAR99\u003c\/li\u003e\n \u003cli\u003ePackage Type: Board Mount\u003c\/li\u003e\n \u003cli\u003eDatasheet: \u003ca href=\"https:\/\/www.waldomapac.com\/upload\/parts\/datasheets\/mf-mr\/mr-xcx15dnc.pdf\" target=\"_blank\"\u003eDownload Datasheet\u003c\/a\u003e\n\u003c\/li\u003e\n\u003c\/ul\u003e\n\n\u003ch3\u003eDisclaimers\u003c\/h3\u003e\n\u003cp\u003eWarning: Do not use these products as safety or emergency stop devices. Failure to follow installation and operation instructions may result in personal injury or death. Specifications are subject to change without notice. For detailed application and installation guidance, refer to the provided datasheet.\u003c\/p\u003e\n\n\u003cp\u003ePart Number: XCX15DNC | Manufacturer: Honeywell | COO: US | UOM: EA | ECCN: EAR99 | Package Type: Board Mount\u003c\/p\u003e\u003cbr\u003e","published_at":"2025-08-16T10:01:47+05:30","created_at":"2025-08-16T10:01:47+05:30","vendor":"Waldom Electonics","type":"Sensors\/Transducers","tags":["Board","HWLL","IMPORT_CSV","IMPORT_PT2","Sensors","Transducers","WLDM"],"price":1408639,"price_min":1408639,"price_max":1408639,"available":true,"price_varies":false,"compare_at_price":null,"compare_at_price_min":0,"compare_at_price_max":0,"compare_at_price_varies":false,"variants":[{"id":50627208446269,"title":"Default Title","option1":"Default Title","option2":null,"option3":null,"sku":"XCX15DNC","requires_shipping":true,"taxable":true,"featured_image":null,"available":true,"name":"XCX15DNC","public_title":null,"options":["Default Title"],"price":1408639,"weight":250,"compare_at_price":null,"inventory_management":"shopify","barcode":"XCX15DNC","requires_selling_plan":false,"selling_plan_allocations":[]}],"images":["\/\/thinkrobotics.com\/cdn\/shop\/files\/XCX15DNC_11944414_024cc281.jpg?v=1755318710"],"featured_image":"\/\/thinkrobotics.com\/cdn\/shop\/files\/XCX15DNC_11944414_024cc281.jpg?v=1755318710","options":["Title"],"media":[{"alt":null,"id":45866948329789,"position":1,"preview_image":{"aspect_ratio":1.35,"height":200,"width":270,"src":"\/\/thinkrobotics.com\/cdn\/shop\/files\/XCX15DNC_11944414_024cc281.jpg?v=1755318710"},"aspect_ratio":1.35,"height":200,"media_type":"image","src":"\/\/thinkrobotics.com\/cdn\/shop\/files\/XCX15DNC_11944414_024cc281.jpg?v=1755318710","width":270}],"requires_selling_plan":false,"selling_plan_groups":[],"content":"\u003ch3\u003eHONEYWELL XCX15DNC Board Mount Pressure Sensor - 15 psi Unamplified\u003c\/h3\u003e\n\n\u003cp\u003eIntroducing the Honeywell XCX15DNC, a high-precision silicon micromachined pressure sensor designed for accurate pressure measurement in various industrial and medical applications. Manufactured in the USA, this board-mounted transducer offers reliable performance with temperature compensation and calibration integrated into an industry-standard package.\u003c\/p\u003e\n\n\u003ch3\u003eProduct Features\u003c\/h3\u003e\n\u003cul\u003e\n \u003cli\u003ePressure range up to 15 psi (1 psi to 240 psi options available)\u003c\/li\u003e\n \u003cli\u003eUnamplified ratiometric millivolt (mV) output\u003c\/li\u003e\n \u003cli\u003eTemperature compensated from 0 °C to 70 °C (32 °F to 158 °F)\u003c\/li\u003e\n \u003cli\u003eAvailable in High Grade (H) for high-performance sensing or Commercial Grade (C) for lower performance needs\u003c\/li\u003e\n \u003cli\u003eProtected against torque-induced stress with a unique stress isolating design\u003c\/li\u003e\n \u003cli\u003eCalibrated offset with high long-term stability and accuracy\u003c\/li\u003e\n \u003cli\u003eSuitable for gage, differential, and absolute pressure measurements\u003c\/li\u003e\n\u003c\/ul\u003e\n\n\u003ch3\u003eElectrical Specifications (at 12 Vdc excitation, 25 °C)\u003c\/h3\u003e\n\u003ctable\u003e\n \u003ctr\u003e\n \u003cth\u003eCharacteristic\u003c\/th\u003e\n \u003cth\u003eXCX\/XCXL C Grade\u003c\/th\u003e\n \u003cth\u003eXCX\/XCXL H Grade\u003c\/th\u003e\n \u003cth\u003eUnit\u003c\/th\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eExcitation voltage\u003c\/td\u003e\n \u003ctd\u003e3.0 - 16.0 Vdc\u003c\/td\u003e\n \u003ctd\u003e3.0 - 16.0 Vdc\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eNull\u003c\/td\u003e\n \u003ctd\u003e±0.5 - ±1.0 mV\u003c\/td\u003e\n \u003ctd\u003e±0.5 - ±1.0 mV\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eTemperature error on offset (0 °C to 70 °C)\u003c\/td\u003e\n \u003ctd\u003e±0.5 mV\u003c\/td\u003e\n \u003ctd\u003e±0.5 mV\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eTemperature error on span (0 °C to 70 °C)\u003c\/td\u003e\n \u003ctd\u003e1.0% span\u003c\/td\u003e\n \u003ctd\u003e1.0% span\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eLinearity, hysteresis error\u003c\/td\u003e\n \u003ctd\u003e0.5 - 1.0% span\u003c\/td\u003e\n \u003ctd\u003e0.3 - 0.5% span\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003eRepeatability\u003c\/td\u003e\n \u003ctd\u003e0.1% span\u003c\/td\u003e\n \u003ctd\u003e0.1% span\u003c\/td\u003e\n \u003ctd\u003e\u003c\/td\u003e\n \u003c\/tr\u003e\n\u003c\/table\u003e\n\n\u003ch3\u003ePressure Range and Output\u003c\/h3\u003e\n\u003ctable\u003e\n \u003ctr\u003e\n \u003cth\u003eFull Scale Pressure\u003c\/th\u003e\n \u003cth\u003eOutput Full Scale Span (mV)\u003c\/th\u003e\n \u003cth\u003eMax Overpressure (psi)\u003c\/th\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e4 in H2O\u003c\/td\u003e\n \u003ctd\u003e38 - 42\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e10 in H2O\u003c\/td\u003e\n \u003ctd\u003e19 - 21\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e0.3 psi\u003c\/td\u003e\n \u003ctd\u003e19 - 21\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e1 psi\u003c\/td\u003e\n \u003ctd\u003e17 - 19\u003c\/td\u003e\n \u003ctd\u003e5\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e5 psi\u003c\/td\u003e\n \u003ctd\u003e57 - 63\u003c\/td\u003e\n \u003ctd\u003e15\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e15 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e45\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e30 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e90\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e60 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e180\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e100 psi\u003c\/td\u003e\n \u003ctd\u003e95 - 105\u003c\/td\u003e\n \u003ctd\u003e200\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e150 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e300\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e240 psi\u003c\/td\u003e\n \u003ctd\u003e95 - 105\u003c\/td\u003e\n \u003ctd\u003e300\u003c\/td\u003e\n \u003c\/tr\u003e\n\u003c\/table\u003e\n\n\u003ch3\u003eAbsolute Output (at 12 Vdc, 25 °C)\u003c\/h3\u003e\n\u003ctable\u003e\n \u003ctr\u003e\n \u003cth\u003eFS Pressure\u003c\/th\u003e\n \u003cth\u003eOutput (mV)\u003c\/th\u003e\n \u003cth\u003eMax Overpressure (psi)\u003c\/th\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e5 psi\u003c\/td\u003e\n \u003ctd\u003e57 - 63\u003c\/td\u003e\n \u003ctd\u003e15\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e15 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e45\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e30 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e90\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e60 psi\u003c\/td\u003e\n \u003ctd\u003e85 - 95\u003c\/td\u003e\n \u003ctd\u003e180\u003c\/td\u003e\n \u003c\/tr\u003e\n \u003ctr\u003e\n \u003ctd\u003e100 psi\u003c\/td\u003e\n \u003ctd\u003e95 - 105\u003c\/td\u003e\n \u003ctd\u003e200\u003c\/td\u003e\n \u003c\/tr\u003e\n\u003c\/table\u003e\n\n\u003ch3\u003eSpecifications \u0026amp; Additional Information\u003c\/h3\u003e\n\u003cul\u003e\n \u003cli\u003ePressure ranges from 4 in H2O to 240 psi\u003c\/li\u003e\n \u003cli\u003eCalibrated for full scale span with ±1.0% (H Grade) or ±2% (C Grade) accuracy over the temperature range\u003c\/li\u003e\n \u003cli\u003eSuitable for medical, vacuum, positive pressure, and small size applications\u003c\/li\u003e\n \u003cli\u003eRoHS compliant\u003c\/li\u003e\n \u003cli\u003eECCN: EAR99\u003c\/li\u003e\n \u003cli\u003ePackage Type: Board Mount\u003c\/li\u003e\n \u003cli\u003eDatasheet: \u003ca href=\"https:\/\/www.waldomapac.com\/upload\/parts\/datasheets\/mf-mr\/mr-xcx15dnc.pdf\" target=\"_blank\"\u003eDownload Datasheet\u003c\/a\u003e\n\u003c\/li\u003e\n\u003c\/ul\u003e\n\n\u003ch3\u003eDisclaimers\u003c\/h3\u003e\n\u003cp\u003eWarning: Do not use these products as safety or emergency stop devices. Failure to follow installation and operation instructions may result in personal injury or death. Specifications are subject to change without notice. For detailed application and installation guidance, refer to the provided datasheet.\u003c\/p\u003e\n\n\u003cp\u003ePart Number: XCX15DNC | Manufacturer: Honeywell | COO: US | UOM: EA | ECCN: EAR99 | Package Type: Board Mount\u003c\/p\u003e\u003cbr\u003e"}
{
"product_id": 10028359090493,
"variant_id" : 50627208446269,
"variant_available" : true,
"inventory_quantity" : 123,
"inventory_management" : "shopify",
"inventory_policy" : "deny",
"image": "//thinkrobotics.com/cdn/shop/files/XCX15DNC_11944414_024cc281.jpg?v=1755318710",
"price": "1408639",
"title": "Default Title",
"value": "Default Title"
}
{
"product_id": "MTAwMjgzNTkwOTA0OTM=",
"variant_id" : "NTA2MjcyMDg0NDYyNjk=",
"variant_available" : "dHJ1ZQ==",
"inventory_quantity" : "MTIz",
"inventory_management" :"c2hvcGlmeQ==",
"inventory_policy" : "ZGVueQ==",
"image":"JTJGJTJGdGhpbmtyb2JvdGljcy5jb20lMkZjZG4lMkZzaG9wJTJGZmlsZXMlMkZYQ1gxNUROQ18xMTk0NDQxNF8wMjRjYzI4MS5qcGclM0Z2JTNEMTc1NTMxODcxMA==",
"price": "MTQwODYzOQ==",
"title": "RGVmYXVsdCtUaXRsZQ==",
"value": "RGVmYXVsdCtUaXRsZQ==",
"variant_options": {"Title": "RGVmYXVsdCtUaXRsZQ=="
}
}
India
---
Australia
Canada
India
Mexico
United Arab Emirates
United States